Impurity Peaks on an HPLC Chromatogram: What They Are and What a Certificate Can't Tell You

909 Aminos LLC · Published 2 November 2026 · 4 min read

On an HPLC purity certificate, everything other than the main peak is reported only as "other integrated peak area". A standard certificate doesn't identify what those minor peaks are.

In short
  1. "99.22% (area %)" means the main peak made up 99.22% of the peak area the method detected and integrated.
  2. The remaining 0.78% is other integrated peak area; the certificate doesn't identify it.
  3. Laboratories can calculate area % on different bases, so figures from different laboratories aren't directly interchangeable.
  4. An HPLC purity figure doesn't cover co-eluting or non-absorbing substances, the amount in the vial, or tests that weren't performed.

1Reading the chromatogram

main peakminor peaksminor peakretention time →detector response
Schematic, not to scale. The tallest peak is normally the target peptide; smaller peaks before or after it are other components that absorb at the method's wavelength.
Horizontal axis

Retention time: when each component left the column.

Vertical axis

Detector response at the method's wavelength.

Tallest peak

Normally the target peptide. On the certificate for lot TESA-2609A, the main peak elutes at 9.249 minutes.

Smaller peaks

Before or after it: other components that absorb at that wavelength.

Example, lot TESA-2609A (KMD certificate 9AL-20260921-C7PG-01):

99.22%Main peak (area %)
0.78%Other integrated peak area
9.249 minMain peak elution
Not identifiedWhat the other peaks are

2What minor peaks can be

General background

This is general background, not an identification of the peaks on any 909 certificate.

In synthetic peptides, minor peaks are often peptide-related substances, molecules very close to the target. Common types include:

TypeWhat it is
Deletion sequenceOne amino acid missing from the chain
Truncated sequenceSynthesis stopped early, leaving a shorter chain
Incompletely deprotectedA protecting group from synthesis still attached
Oxidized formTypically at methionine, tryptophan or cysteine
Deamidated formAsparagine or glutamine converted to their acid forms
DiastereomerOne amino acid in the opposite (D/L) configuration
What the certificate doesn't establish

A standard purity certificate reports the minor peaks only as a share of total area. It doesn't name them, and it doesn't establish which of these types, if any, are present. Identifying individual peaks takes additional analysis, usually mass spectrometry.

3Two laboratories, two definitions

Purity by area % depends on what goes into the total:

KMD Analytical (lot TESA-2609A)ILS Laboratories (lot KLOW-2610A)
Result99.22%99.89%
Stated basisPercentage of total integrated peak areaTarget peptide(s) relative to all peptide-related peaks; non-peptide process-related impurities, if detected, excluded
DetectionHPLC-DADRP-HPLC at 214 nm
KMD Analytical

Divides by the total integrated peak area.

ILS Laboratories

Divides by peptide-related peaks only.

The takeaway

Figures calculated on different bases aren't directly interchangeable, so read the stated basis before comparing purity results from different laboratories.

4What an HPLC purity figure can't tell you

Co-eluting components

A substance that leaves the column at the same time as the main peak is counted as part of it.

Different detector response

Area is detector response, not mass, so a component that absorbs weakly is under-represented and one that absorbs strongly is over-represented.

Non-absorbing substances

Water, many salts and some solvents give little or no signal at typical wavelengths.

Amount in the vial

Purity is a proportion. The quantity of peptide is a separate content result. See Purity vs. Measured Content.

Tests not performed

KMD's certificates state that, unless expressly noted, no assessment was made of microbial, endotoxin, heavy-metal or residual-solvent content. A certificate covers only the tests it lists.

5Checklist

?Frequently asked questions

Does 99% purity mean 99% of the powder is the peptide?

No. It means the main peak was 99% of the detected peak area. How much of the powder is peptide is a content question, affected by counter-ions and water.

Can a certificate tell me what each small peak is?

Not a standard HPLC purity certificate. It reports minor peaks only as a share of area; identifying them needs further analysis such as mass spectrometry.

Why do purity figures from different laboratories differ?

Laboratories can use different columns, wavelengths and definitions of the total. Compare figures only when the basis is the same.