An identity test assesses whether a sample's main component corresponds to the compound named on the label; how strong that evidence is depends on the method used.
An identity test answers a single question: is the main component of the sample the compound named on the label? Certificates answer it in different ways, and the method matters as much as the word "Conforms". This page explains the two approaches you are most likely to see, what each one can and can't establish, and how to read the identity line on a 909 certificate.
Three separate questions appear on a typical certificate:
| Question | Typical result | Covered in |
|---|---|---|
| Is it the right compound? | Identity: conforms / does not conform | This page |
| How much of what was detected is that compound? | Purity, area % | Purity vs. net peptide content |
| How much of it is in the vial? | Content, mg per vial | Purity vs. net peptide content |
A high purity figure doesn't prove identity on its own. A method can report that one component dominates the sample without establishing what that component is.
This approach uses the same HPLC run that produces the purity result. A reference standard of the expected compound is analyzed under the same conditions, and the sample is compared with it on two properties.
Retention time (RT). The time a component takes to travel through the column and reach the detector. Under fixed conditions (column, mobile phase, gradient, temperature, flow rate) a given compound elutes at a reproducible time. A sample peak that elutes where the reference elutes is consistent with the same compound.
UV spectrum. A diode-array detector (DAD) records the full ultraviolet absorbance spectrum across each peak, not just a single wavelength. The spectrum's shape depends on the parts of the molecule that absorb UV light. In peptides, that's mainly the peptide backbone and aromatic amino acids such as tryptophan, tyrosine and phenylalanine. A sample spectrum that matches the reference spectrum adds a second, independent point of agreement.
What a match provides: supporting evidence of identity. The main peak behaves like the reference standard on two properties, under that laboratory's method. Agreement on retention time and UV spectrum is consistent with the expected compound, but it is not a definitive molecular identification.
What it doesn't establish:
Mass spectrometry (MS) ionizes the sample and measures the mass-to-charge ratio of the ions. Peptides usually carry several charges, so the spectrum shows a series of ions from which the molecular mass is calculated. Identity is assessed by comparing the observed mass with the theoretical mass of the expected sequence, within a stated tolerance. Liquid chromatography is often coupled directly to the mass spectrometer (LC-MS), so each peak's mass is measured as it elutes.
What MS adds:
Its limits:
| RT + UV vs. reference | Mass spectrometry | |
|---|---|---|
| Measures | Elution time and UV spectrum shape | Mass-to-charge ratio, from which mass is calculated |
| Needs a reference standard | Yes | Not for intact mass (compared with the theoretical value) |
| Confirms molecular mass | No | Yes |
| Separates same-mass variants | Sometimes, by elution | No, without fragmentation (MS/MS) |
| Same run as the purity test | Yes | Often a separate run or instrument |
| Quantitative | Through the HPLC peak areas | Not directly |
The methods are orthogonal: they rely on different physical properties, so agreement between them is stronger evidence than either one alone.
The certificate for lot TESA-2609A (KMD Analytical, certificate 9AL-20260921-C7PG-01) reports:
| Field | As stated |
|---|---|
| Identity | Identity Conforms (RT + UV vs. reference) |
| Main peak retention time | 9.249 minutes |
| Method | HPLC-DAD |
Read accurately, this means the main peak's retention time and UV spectrum matched the laboratory's reference under the laboratory's HPLC-DAD method. This certificate does not include a mass-spectrometry result: the laboratory's published scope is identity, purity and content analysis by HPLC-DAD. So the certificate provides supporting identity evidence (chromatographic and spectral agreement with a reference), not a definitive molecular identification, and it doesn't report a measured molecular mass.
No. Matching retention time and UV spectrum against a reference standard is supporting evidence of identity under that laboratory's method. It does not measure molecular mass or confirm the exact sequence.
A direct measurement of molecular mass, compared with the theoretical mass of the expected sequence. It can separate many related impurities that differ in mass, but not same-mass variants without fragmentation (MS/MS).
No. It reports identity by retention time and UV spectrum against a reference, using HPLC-DAD. The laboratory's published scope is identity, purity and content analysis by HPLC-DAD.
No. A retention time is meaningful only under the conditions of the method that produced it, such as the column, mobile phase, gradient, temperature and flow rate.
Sources are cited for reference. Their publishers have not reviewed or endorsed this guide.